Effect of different surface finishing and of hydroxyapatite coatings on passive and corrosion current of Ti6Al4V alloy in simulated physiological solution

Academic Article
Publication Date:
1997
abstract:
Direct and alternating current electrochemical tests were carried out on Ti6Al4V with different surface finishing and with hydroxyapatite (HA) coatings. Sand-blasting and rough titanium deposits obtained by vacuum plasma spraying (VPS) bring about an increase of passive and corrosion current density (c.d.) with respect to smooth Ti6Al4V, as a consequence of the augmentation of the real surface. The presence of HA deposits obtained by VPS causes an increase of passive and corrosion c.d. of the metallic substrate of about one order of magnitude and this should be taken into account in view of human body applications.
Iris type:
1.1.01 Articoli/Saggi in rivista - Journal Articles/Essays
List of contributors:
Cabrini, Marina; Cigada, Alberto; Rondelli, G.; Vicentini, B.
Authors of the University:
CABRINI Marina
Handle:
https://aisberg.unibg.it/handle/10446/82672
Published in:
BIOMATERIALS
Journal
  • Research

Research

Concepts


Settore ING-IND/22 - Scienza e Tecnologia dei Materiali